AOIP-advanced field documenting thermometer / temperature calibrator (THERMYS 150)

2
544
Vui lòng liên hệ với chúng tôi– chúng tôi sẽ liên hệ lại với khách hàng bằng điện thoại hoặc email.
Lưu ý: Tham khảo ý kiến của nhân viên INO sẽ giúp bạn tiết kiệm được thời gian và chi phí khi cần mua sắm. ​​Với sự tư vấn của chúng tôi, bạn sẽ không gặp khó khăn khi tìm hiểu về đặc tính của sản phẩm cần mua.

THERMYS 150 is an all-in-one high accurate documenting thermometer / temperature calibrator for thermocouples (14 types), resistive probes (12 types) and thermistors with a high accuracy:

  • Thermocouples: Up to 0.005% RDG
  • RTDs: Up to 0.006% RDG
  • Thermistors: Up to 0.006% RDG 

It is designed for field and lab use for advanced temperature maintenance and use on test bench in all industries.

  • Simultaneous measurement and emission of temperature (IN / OUT)
  • Simultaneous measurement over two channels for comparison calibration (IN / IN)

The second channel can be used either for temperature generation or measurement. When used as a dual input thermometer, THERMYS 150 is the perfect tool for temperature comparison calibration and differential measurement. When the second channel is used as an output, the instrument can be used to control temperature recorders, indicators, handheld thermometers..

Using this user-friendly instrument, calibration tasks can be quickly carried out over the whole process chain. Take the 900 g documenting process calibrator to the field with you during the whole week with 10 calibration procedures stored in the device. Run the procedure after connecting the probes to the instrument (Easy connect system®) and save the results for onsite easy and quick calibration. Back to the office, you can then upload the data on a computer in order to issue customized calibration certificates with dedicated calibration software DATACAL.

 Providing extended functionalities (temperature simulation, scaling, steps, synthesizer, statistical functions…) and audit trails, THERMYS 150 complies with both 21 CFR Part 11 and NADCAP Heat Treatment standards and makes advanced data exploitation and full data traceability easier. 

IP 54, fully protected by an antichoc rubber holster, THERMYS 150 integrates "easyconnect" terminals and a wide backlite display that makes it easy to use in any severe or dark conditions.

THERMYS 150 has also the capability to drive baths and dry-blocks when associated with the specific cable (ref. ACL600).

Specifications and performances in temperature @23°C ±5°C

Uncertainty is given in % of reading (THERMYS 150 display) + fixed value.

Resistive probes: Measurement and simulation

Sensor Range (Input and Output) Resolution Accuracy / 1 year (Measurement) Accuracy / 1 year (Simulation)
Pt50 (= 3851) -220°C to +850°C 0.01°C 0.006% RDG + 0.04°C 0.006% RDG + 0.04°C
Pt100 (= 3851) -220°C to +850°C 0.01°C 0.006% RDG + 0.03°C 0.006% RDG + 0.03°C
Pt100 (= 3916) -200°C to +510°C 0.01°C 0.006% RDG + 0.03°C 0.006% RDG + 0.03°C
Pt100 (= 3926) -210°C to +850°C 0.01°C 0.006% RDG + 0.03°C 0.006% RDG + 0.03°C
Pt200 (= 3851) -220°C to +850°C 0.01°C 0.006% RDG + 0.04°C 0.006% RDG + 0.04°C
Pt500 (= 3851) -220°C to +850°C 0.01°C 0.006% RDG + 0.03°C 0.006% RDG + 0.03°C
Pt1000 (= 3851) -220°C to +850°C 0.01°C 0.006% RDG + 0.03°C 0.006% RDG + 0.03°C
Ni100 (= 618) -60°C to 180°C 0.01°C 0.006% RDG + 0.05°C 0.006% RDG + 0.05°C
Ni120 (= 672) -40°C to +205°C 0.01°C 0.006% RDG + 0.05°C 0.006% RDG + 0.05°C
Ni1000 (= 618) -60°C to +180°C 0.01°C 0.006% RDG + 0.05°C 0.006% RDG + 0.05°C
Cu10 (= 427) -50°C to 150°C 0.10°C 0.006% RDG + 0.18°C 0.006% RDG + 0.18°C
Cu50 (= 428) -50°C to +150°C 0.01°C 0.006% RDG + 0.05°C 0.006% RDG + 0.05°C

Resistive probes measurements in 2, 3 or 4 wires: automatic recognition of number of connected wires, with indication on screen
Accuracies are given for 4-wire mounted probes
Take into account particular error of temperature sensor used and implementation conditions
Admissible measuring current: 0.01 mA to 1 mA
In simulation mode, specifications given for 1 mA measuring current (Pt50 / 100, Ni100 / 120, Cu10 / 50) or 0.1 mA (Pt200 / 500 / 1000, Ni1000)
Establishing time:  <1ms for simulation (simulation on quick transmitters)
Temperature coefficient: < 10% of accuracy /°C

Resource & Download

Lưu ý: Nếu một thiết bị nào đó không được liệt kê ở đây, điều đó không có nghĩa rằng chúng tôi không hỗ trợ được bạn về thiết bị đó. Hãy liên hệ với chúng tôi để biết danh sách đầy đủ về thiết bị mà chúng tôi có thể hỗ trợ và cung cấp.
INO: Bán, Báo giá, tư vấn mua sắm và cung cấp, tư vấn sản phẩm thay thế; tương đương, hướng dẫn sử dụng, giá…VNĐ, …USD info@ino.com.vn | 086 9421303 | 02873000184 | AOIP-advanced field documenting thermometer / temperature calibrator (THERMYS 150).

Comments are closed.