SA32 data acquisition systems are dedicated to measurement, monitoring and recording of analogue signals on 32 channels in 2 wires or 8 channels in 4 wires.
Designed with flexibility and precision, this system can be used in numerous applications, on site or in laboratory.
SA 32 allows 100 addresses to be programmed, which can be defined to measure real channels (up to 32 channels) or virtual channels used to perform calculations / maths functions.
Input channels are scanned by a CMOS logic circuit.
Scanning is organised in «tasks» or scanning procedures, defining what to scan (channels), how to scan (scanning start and stop, scanning rate), and how to process the results (mass storage, RS 232 output). The user has access to 16 tasks, therefore 16 different scanning subgroups.
The datalogger is completely programmable via keyboard and help menu on screen, or via RS 232 interface using LTC 32W software supplied with the unit.
Backlit LCD display shows 2 channels simultaneously either under digital or graphical format (150 points definition) on a time base adjustable between 1 s to 1 min. It also enables stored valued to be recalled on screen for display under digital or graphic format.
SA 32 perform measurement, monitoring and recording of analogue signals coming from sensors of physical or electrical values. These signals can be:
- Voltage: 0-10 V
- Current: 0-20 and 4-20 mA with external shunts
- Thermocouples: Type K/T/J/S/B/N with or without cold junction compensation
- Resistance: 3000 Ω
- RTD: Pt100 in 3 or 4 wires
- Dry contacts
Specifications and performances in temperature @23°C ±1°C
Uncertainty is given in % of reading + fixed value.
Resistive probes: Measurement
Type | Range | Resolution 7 samples / s |
Resolution 20 samples / s |
Resolution 50 samples / s |
Accuracy / 1 year 7 samples / s |
Accuracy / 1 year 20 samples / s |
Accuracy / 1 year 50 samples / s |
Pt100 | -220°C to +850°C | 0.01°C | 0.1°C | 1°C | 0.025% RDG + 0.1°C | 0.025% RDG + 0.15°C | 0.04% RDG + 0.6°C |
Accuracies are given for 4-wire mounted probes
Admissible measuring current: 1 mA
Admissible line resistance: 100 Ω per wire
Temperature coefficient: < 0.003% /°C
Thermocouples: Measurement
Type | Range | Resolution 7 samples / s |
Resolution 20 samples / s |
Resolution 50 samples / s |
Accuracy / 1 year 7 samples / s |
Accuracy / 1 year 20 samples / s |
K | -250 to -200°C -200 to -100°C -100 to +1370°C |
0.5°C 0.2°C 0.1°C |
0.5°C 0.2°C 0.1°C |
1°C 1°C |
0.01% RDG + 1°C 0.01% RDG + 0.4°C 0.01% RDG + 0.2°C |
0.02% RDG + 1.4°C 0.02% RDG + 0.5°C 0.02% RDG + 0.3°C |
T | -250 to -200°C -200 to -100°C -100 to +400°C |
0.5°C 0.2°C 0.1°C |
0.5°C 0.2°C 0.1°C |
1°C 1°C |
0.01% RDG + 0.8°C 0.01% RDG + 0.4°C 0.01% RDG + 0.2°C |
0.02% RDG + 1°C 0.02% RDG + 0.5°C 0.02% RDG + 0.3°C |
J | -210 to -120°C -120 to1100°C |
0.2°C 0.1°C |
0.2°C 0.1°C |
1°C 1°C 1°C |
0.01% RDG + 0.3°C 0.01% RDG + 0.2°C |
0.02% RDG + 0.4°C 0.02% RDG + 0.3°C |
S | -50 to +120°C +120 to +1768°C |
1°C 1°C |
1°C 1°C |
1°C 1°C |
0.01% RDG + 1.2°C 0.01% RDG + 0.7°C |
0.02% RDG + 1.7°C 0.02% RDG + 1°C |
B | +400 to +900°C +900 to +1820°C |
1°C 1°C |
1°C 1°C |
1°C 1°C |
0.01% RDG + 1.5°C 0.01% RDG + 0.8°C |
0.02% RDG + 2°C 0.02% RDG + 1°C |
N | -250 to -200°C -200 to -120°C -120 to +1300°C |
1°C 0.2°C 0.1°C |
1°C 0.2°C 0.1°C |
1°C 1°C 1°C |
0.01% RDG + 1.5°C 0.01% RDG + 0.6°C 0.01% RDG + 0.3°C |
0.02% RDG + 2°C 0.02% RDG + 0.9°C 0.02% RDG + 0.4°C |
Accuracy is given for reference @ 0°C.
When using the internal reference junction, add an additional uncertainty at 0°C of ±0.5 °C in 2-wire measurement and ±1°C in 1-wire measurement.
Input impedance: > 1000 MΩ
Measuring current: < 5 nA (typical: 300 pA)
Admissible line resistance: 100 Ω per wire
Temperature coefficient: < 10% of accuracy /°C
Specifications and performances in process @23°C ±1°C
DC voltage: Measurement
Range | Resolution 7 samples / s |
Resolution 20 samples / s |
Resolution 50 samples / s |
Accuracy / 1 year 7 samples / s |
Accuracy / 1 year 20 samples / s |
60 mV | 1 µV | 1 µV | 20 µV | 0.02% RDG + 6 µV | 0.02% RDG + 6 µV |
600 mV | 2 µV | 10 µV | 100 µV | 0.02% RDG + 10 µV | 0.02% RDG + 30 µV |
6 V | 20 µV | 100 µV | 1 mV | 0.02% RDG + 0.1 mV | 0.02% RDG + 0.3 mV |
0-10 V | 20 µV | 100 µV | 1 mV | 0.02% RDG + 0.1 mV | 0.02% RDG + 0.3 mV |
Input impedance: > 1000 MΩ
Measuring current: < 5 nA (typical: 300 pA)
Acceptable maximum voltage in common mode: 50 V- or ~ peak
Acceptable maximum voltage between different inputs: 50 V- or ~ peak
Temperature coefficient: < 10% of accuracy /°C
DC current: Measurement
With or without loop supply
Range | Measuring range | Accuracy / 1 year 7 samples / s |
Accuracy / 1 year 20 samples / s |
0-20 mA | 0 mA to 20 mA | 0.1% RDG | 0.1% RDG |
4-20 mA | 4 mA to 20 mA | 0.1% RDG | 0.1% RDG |
With 50 Ω external shunt
Resistance: Measurement
Range | Resolution 7 samples / s |
Resolution 20 samples / s |
Resolution 50 samples / s |
Accuracy / 1 year 7 samples / s |
Accuracy / 1 year 20 samples / s |
60 Ω | 1 mΩ | 1 mΩ | 10 mΩ | 0.025% RDG + 6 mΩ | 0.025% RDG + 6 mΩ |
600 Ω | 2 mΩ | 10 mΩ | 100 mΩ | 0.025% RDG + 10 mΩ | 0.025% RDG + 30 mΩ |
3000 Ω | 20 mΩ | 100 mΩ | 1 Ω | 0.025% RDG + 0.1 Ω | 0.025% RDG + 0.3 Ω |
Uncertainties given for 4-wire measurement
Measuring current: 1 mA
Admissible line resistance: 100 Ω per wire
Dry contacts: Measurement
Measuring current: 1 mA
Level 0: line resistance R0 ³ 100 kΩ
Level 1: line resistance R1 £ 1 kΩ
Relay outputs
Relay breaking capacity on resistance load: 48 V, 1 A or 30 Ω
Switching life: 5.105
Relays are triggered on alarm presence, conditional signal or command from the communication interface.
Analogue output
Range : 0 – 2.55V
Measureing range : 10mV
Accuracy / 1 year : 10mV
Comments are closed.