P200 PRO is a field calibrator that measures and generate pressure from -1 to 700 bars with an accuracy of 0.025% FS. It is developed for the calibration, verification and testing of pressure sensors and transmitters, manometers and pressostats.
Besides, it measures and generates process voltage and current and has a Pt100 input for calibrating flow transmitters.
- Pressure range: -1 to 700 bars
- Leak test
- Switch test
- Calibration of transmitters and pressostats
P200 PRO is a versatile, robust and handy instrument with its IP65 measurement head and removable pump. It is a comprehensive calibration system yet easy-to-use. It has an internal pressure sensor and accepts an external module in addition.
It simultaneously displays the output pressure and input current or voltage, the pressure percentage, the pressostat status or the MIN / MAX pressure value.
P200 PRO stores not only 1,500 measurement values but several calibration procedures, leak test or switch test procedures.
The instrument is delivered in standard with a factory calibration report traceable to the national standards that states the nominal and actual values and deviation errors.
The pressure calibrator is widely used and recommended for maintenance, regulation, metrology departments of process industries, energy field, applied or scientific research activities.
Specifications and performances @23°C ±5°C
Pressure
Pressure media: AISI 316 SS compatible fluids (water, gas, and oil)
Engineering units: mbar, bar, hPa, kPa, Mpa, kg/cm2, kg/m2, psi, mmH2O, cmH2O, mH2O, Torr, atm, lb/ft2, inH2O, FTH2O, mmHg, cmHg, mHg, inHg
Accuracy: accuracies given for 1 year, include non linearity, hysteresis, and repeatability.
Temperature coefficient inside the temperature compensated range: ±0.002% RDG /°C
Temperature compensation: Automatic with built-in calibration matrix from 0°C to 45°C
By interne moduel | By external sensor | |
Accuracy | ±0.025% FS | ±0.025% FS |
Overpressure | 125% FS | 125% FS |
Port | 3/8" BSP male | 1/4" BSP male |
Connexion cable | Internal | 2 m |
Process voltage
Type | Measurement / Emission | Range | Resolution | Accuracy / 1 year |
Voltage | IN / OUT | -200 to 200 mV | 1 mV | 0.02% RDG + 3 mV |
-0.2 to 2 V | 10 mV | 0.02% RDG + 10 mV | ||
-2 to 20 V | 100 mV | 0.02% RDG + 100 mV |
Process current
Type | Measurement / Emission | Range | Resolution | Accuracy / 1 year |
Current | IN / OUT | -5 to 22 mA | 0.1 mA | 0.02% RDG + 0.4 mA |
Loop supply: 24 V ±10% (30 mA maximum)
Temperature
Type | Measurement / Emission | Range | Resolution | Accuracy / 1 year |
Pt100 (=3926) | Measurement only | -200°C to +850°C | 0.01°C | 0.02% RDG + 0.05°C |
Further features
Scaling in measurement mode | Scaling allows process signals to be displayed in % of FS or in pressure, offering a direct comparison between the measured pressure value and the transmitter output value and allowing sensors to be corrected after a calibration. |
Transmitters tests | The feature enables any pressure transmitter to be controlled and calibrated with simultaneous display of input and output values in % F.S. or in actual unit. The measuring circuit is also able to power the loop for a direct connection to the transmitter under test. |
Switch test | In pressure, P200 PRO controls and displays the applied pressure and the switch status OPEN or CLOSED of pressostats. |
Leak test | P200 PRO detects the presence of any pressure leakage through a test procedure carrying out pressure fall measurements in a programmable time interval. |
General specifications
Size | 130 x 120 x 70 mm |
Weight | Nett: 1.4 kg Gross: 2.5 kg |
Display | Liquid crystal graphical display with automatic and manual backlite |
Measurement sampling time | 250 ms |
Battery | Type: Rechargeable Ni-Mh with external charger Charging time: 5 h at 90% and 6 h at 99%, if instrument switched off Life time: 10 h in measurement mode without backlight |
Communication ports | RS 232, full bidirectional |
Storage capacity | 1,500 measured values with date and time stamp |
Environmental specifications
Reference range | 23°C ±5°C |
Operating reference range | -10 to 55°C (RH: 95 % non condensing) |
Storage temperature limits | -30°C to +60°C (without battery) |
IP protection | IP65 according to EN60529 |