Vui lòng liên hệ với chúng tôi– chúng tôi sẽ liên hệ lại với khách hàng bằng điện thoại hoặc email.
Lưu ý: Tham khảo ý kiến của nhân viên INO sẽ giúp bạn tiết kiệm được thời gian và chi phí khi cần mua sắm. Với sự tư vấn của chúng tôi, bạn sẽ không gặp khó khăn khi tìm hiểu về đặc tính của sản phẩm cần mua.
TAS7000 3D Imaging Analysis System
The TAS7000(∗1) 3D(∗2) imaging analysis system exploits the unique properties of terahertz waves(∗3).The TAS7000 enables 3D non-destructive analysis of the internal structure and chemical composition of the target object; the results are shown on a 3D display. In addition, this system enables measurement of spectral characteristics and 2D(∗4) mapping.
Advantest has developed an unprecedented solution for 3D analysis of industrial materials such as plastics and ceramics, as well as pharmaceutical products.
- A global first: non-destructive 2D and 3D analysis utilizing Terahertz wave
The TAS7000 enables non-destructive, 2D and 3D analysis of the internal composition and structure of measurement targets, utilizing computed tomography (CT) and exploiting the unique properties of terahertz waves.
The TAS7000 enables not only analysis of the structural homogeneity of an internal substance, but also identification of constituents and quantitative distribution measurement.
Moreover, in contrast with other regions of the electromagnetic spectrum, terahertz waves can penetrate various types of materials with moderate transmittance, allowing for analysis. - High-speed measurement utilizes Advantest's terahertz optical sampling technology
The TAS7000 relies on Advantest's newly developed, ultra-short pulse femtosecond optical fiber laser technology for terahertz wave generation and detection, and utilizes the company's original terahertz wave high-speed optical sampling technique. Thus, it delivers measurement throughput approximately 1000 times greater than the previous system. - Broadband terahertz wave spectroscopic analysis up to 3THz
The TAS7000 enables broadband, high-resolution spectroscopic analysis up to a maximum analytical frequency of 3THz. The system is ideally suited for acquiring fingerprint spectra of a comprehensive range of chemical, industrial, and biological materials. For high precision, the system has the option to remove water vapor from the sample chamber by flowing dry air during the measurement.
Applications | Non-destructive analysis, 3D constituent density distribution analysis, 3D spectroscopic measurement |
---|---|
Analytical functions | 2D / 3D computed tomography (CT) density distribution analysis Terahertz spectroscopic transmission analysis:
|
Analytical frequency range (∗1) | High dynamic range module : 0.02THz to 0.6THz Broadband module(∗2) : 0.05THz to 3THz |
Dynamic range (∗1) | High dynamic range module : ≥60dB (at peak, 3min.) Broadband module : ≥50dB (at peak, 15min.) |
Spatial resolution of CT measurement function (∗1) | ≤3mm (using wire phantom; modulation transfer function (MTF), 10%) |
Throughput | Maximum waveform acquisition speed : 250Hz (Spectroscopic measurement mode) Acquisition time for CT cross section : ≤15min. (Measurement of 1 CT cross section is divided into 64×64) |
Maximum sample size and weight | ≤φ310mm (12.2inch) × 310mm (12.2inch) (H) ≤20kg |
Operating environment | Temperature : +10°C to +30°C Relative humidity : ≤80% (non-condensing) |
Power requirements | AC100V (100-120) / 200V (220-240), 850VA, 50/60Hz |
Dimensions | Size : 1500mm (W) × 1570mm (D) × 1600mm (H) Mass : 400kg or less (∗3) |