The TAS7400 product line offers a series of low-cost, all-purpose spectroscopy systems that enable spectroscopic measurements using terahertz (THz) waves. The tools are capable of performing non-destructive analysis on a wide variety of sample types, making it applicable to a broad range of fields from life sciences to electronics, where precise chemical and material characterization is critical. The systems are ideal for settings ranging from basic research to product development, as well as for manufacturing and quality control, and employ easy-to-use spectroscopic methods that do not require specialized knowledge of THz wave generation or optics.
- Low-cost, all-purpose THz spectroscopy system offers superior spectroscopy performance
- Compact desktop form factor
- THz spectroscopic analysis in a range of frequencies (0.03–7 THz)
- Transmission, reflectance, ATR (Attenuated Total Reflection), and transmission polarization measurement modes support spectroscopic analysis of a wide variety of materials
- Dry air filter unit attached, so that measurements will be unaffected by water vapor absorption
- A full range of holders and cells to handle aqueous solutions, powders, and other types of samples
Terahertz spectroscopy system
|
||||
---|---|---|---|---|
TAS7400SL (Low-frequency system) |
TAS7400SP (Standard system) |
TAS7400SU (Wideband system) |
||
Measurement mode | Spectroscopy (Transmission/ reflectance) ∗1 |
Spectroscopy (Transmission/ reflectance/ ATR/ transmission polarization) ∗1 |
Spectroscopy (Transmission/ reflectance/ ATR) ∗1 |
|
Analytical targets | Dielectric materials, chemical materials and other reagents |
Chemical materials, construction materials, organisms, medicines, and other reagents | ||
Specimen dimensions | Transmission/ reflectance mode φ20–30 mm, Thickness 10 mm or less |
Transmission/ reflectance mode φ5–30 mm, thickness 10 mm or less ATR mode |
||
Measurement performance |
Frequency range(∗2) |
0.03-2 THz | 0.1–4 THHz | 0.5–7 THz (transmission/ reflectance mode) 0.5–6.5 THz (ATR mode) |
Frequency accuracy(∗2) |
±10 GHz or less (at 0.56 THz) |
±10 GHz or less (at 1.4 THz) |
±10 GHz or less (at 1.4 THz) |
|
Frequency resolution | 1.9 GHz / 7.6 GHz | 1.9 GHz / 7.6 GHz | 1.9 GHz / 7.6 GHz | |
Dynamic range(∗2) (at peak frequency) |
50 dB or higher | 60 dB or higher | 57 dB or higher (transmission/reflectance mode) 55 dB or higher (ATR mode) |
|
Throughput | 200 ms/scan | |||
Measurement accessories | Transmission mode Solid holder, liquid/powder cell∗3, dry air isolation kit∗3, rotating holder∗3 Reflectance mode ATR mode (SP/SU only) |
|||
Display function | Spectral display (transmittance, reflectance, ATR*, phase difference, absorbance, absorption coefficient, complex refractive index complex permittivity) *ATR with SP/SU only Time response display (electric field strength) |
|||
Software(∗3) | Polarization application, automatic control, FDA21CFR Part11 supported, offline analysis | |||
Dry air purge | Dry air unit attached (external dry air supply necessary) | |||
External accessory(∗4) | Temperature control accessory (-10 to +80 °C type, room temp to +300 °C type) | |||
Controller | Included as standard (OS: 64-bit Windows 7 Professional) | |||
Data file format | Proprietary binary, JCAMP-DX, SPC, CSV | |||
General specifications |
Usage Environment | Temperature range: +10 °C to +30 °C Relative humidity: Under 80% (no condensation) |
||
Storage environment | Temperature range: -10 °C to +50 °C Relative humidity: Under 80% (no condensation) |
|||
Power |
Analysis unit: AC100 V (100–120) / 200 V (220–240) ±10%, 50/60 Hz, 180 VA Measurement unit: AC100 V (100–120) / 200 V (220–240) ±10%, 50/60 Hz, 150 VA *Not including analysis PC |
|||
Size/weight | Analysis unit: Dimensions: 430 (W) × 540 (D) × 330 (H) mm / 30 kg or less Measurement unit: Dimensions: 500 (W) × 490 (D) × 410 (H) mm /40 kg or less |
Comments are closed.