Lưu ý: Tham khảo ý kiến của nhân viên INO sẽ giúp bạn tiết kiệm được thời gian và chi phí khi cần mua sắm. Với sự tư vấn của chúng tôi, bạn sẽ không gặp khó khăn khi tìm hiểu về đặc tính của sản phẩm cần mua.
The UPS III is a high accuracy loop calibrator. Its size makes it rugged and extremely compact. Measuring 3” x 5” and weighing just 9.7 ounces, it fits comfortably in a shirt pocket.
It is an essential tool for loop testing, instrument maintenance and valve set-up, with an easy to read display and simple to use time saving features.
Features
- Measure or Source 0 to 24 mA
- Accuracy 0.01% of reading
- Dual mA and % readout, linear or flow
- Step, Span Check, Valve Check, Ramp
- 6 VDC measurement and continuity
- HART® compatible
|
Applications
Graphic Display The graphic display with menu-driven interface is easier to use than traditional knobs, switches, multi-function keys and dual key sequences.
Measure or Source 0 to 24 mA An internal 24 V loop supply is available in both measure and source modes, essential during plant shutdowns.
Accuracy 0.01% of Reading This includes 12 month stability and temperature effects. The UPS III can maintain a 4:1 calibration ratio with the latest instrumentation and is typically 20 to 80 times more accurate than a Digital Multi-Meter (DMM).
Simultaneous % Readout Displays mA and the percentage value of 4 to 20 mA or 0 to 20 mA. The UPS III also converts mA readings to % flow. This is not possible with many loop calibrators.
Step, Span Check and Value Check Step quickly through linearity tests and zero/span adjustments. In valve mode, seating can be checked and the positioner set.
Auto Step and Ramp Outputs have adjustable rates (1 to 599 s) for single-handed loop testing, valve set-up and slew rate checks.
“Fast Set” Output Allows values to be set to within 0.001 mA using the arrow keys. Holding a key quickly ramps the output to the next value. The system is ideal for alarm trip tests.
60 VDC Measurement For loop diagnostics and maintenance of voltage output instruments.
Continuity and Switch Test Facility This reduces DMM dependence when fault finding, checking loop integrity and testing switches.
Hart Compatible Internal 235 O loop resistor (menu selectable)
EMC compliance to EN 61326-1 Complies with heavy industry standard. Most loop calibrators are only suitable for light industrial use.
|
Specifications
Performance
Electrical
Power Supply |
4 x 1.5 V AA-size alkaline batteries (standard) or universal power supply (see option B) |
Battery Life |
75 hours in read mode, 18 hours at 12 mA |
Auto Power Down |
30 minutes after last key press |
Low Battery Warning |
Battery symbol displayed |
Open Loop |
Flashes “open loop” |
Loop Resistance High |
Flashes “check loop O” |
Out of Range |
Displays (over) |
Hart Resistor |
235 O menu selectable loop resistor |
Environmental
Calibration Reference |
72°F (22°C) ±2°F(±3.6°C)/RH 45% ±15% |
Operating Temperature |
14°F to 122°F (-10°C to 50°C) |
Relative Humidity |
0% to 90% non-condensing |
Conformity |
EN 61010, /EN 61326-1:1997 + A1:1998, CE marked |
Physical
Housing Material |
High impact ABS |
Dimensions (w x h x d) |
3 in x 5 in x 1 in (77 mm x 129 mm x 24 mm) |
Weight |
9.7 oz (275 g) including batteries |
Display |
Graphic LCD, 1.2 in x 2.1 in (31 mm x 54 mm) |
Electrical Terminals |
Gold plated, 4 mm sockets |
|
Lưu ý: Nếu một thiết bị nào đó không được liệt kê ở đây, điều đó không có nghĩa rằng chúng tôi không hỗ trợ được bạn về thiết bị đó.
Hãy liên hệ với chúng tôi để biết danh sách đầy đủ về thiết bị mà chúng tôi có thể hỗ trợ và cung cấp.
INO: Bán, Báo giá, tư vấn mua sắm và cung cấp, tư vấn sản phẩm thay thế; tương đương, hướng dẫn sử dụng, giá…VNĐ, …USD info@ino.com.vn | 086 9421303 | 02873000184 | DRUCK-Loop Calibrator ( Model: UPS-III).
Comments are closed.