Lưu ý: Tham khảo ý kiến của nhân viên INO sẽ giúp bạn tiết kiệm được thời gian và chi phí khi cần mua sắm. Với sự tư vấn của chúng tôi, bạn sẽ không gặp khó khăn khi tìm hiểu về đặc tính của sản phẩm cần mua.
The Hybrid 2000 Wound Component Scanner is a production safety tester designed for test requirements of three-phase motors, double-wound transformers and heater related wound components. Quality verification test items for wound components consist of AC/DC Hipot (dielectric withstanding), Insulation Resistance (IR), and Impulse Winding tests. We have integrated these tests into one unit.
- Programmable hipot output voltage to 5kV AC and 6kV DC
- Trip current programmable to 30mA AC and 10mA DC
- Insulation resistance to 50GΩ / 5KV DC
- DCR 10mΩ ~ 100kΩ
- Built-in 8 channel scanner
- Storage of 20 tests setups with 50 steps per setup
Applications
The Hybrid 2000 is a comprehensive tester designed for wound component safety and design verification testing requirements.
- Motor, Fan
- Electric Heater Tube
- Transformer
- Switch, Wire
- Camera Micro Motor, Coil
Specifications
Hybrid 2000 |
Mode |
AC/DC/IR/ DCR -8CH/ IWT |
Impulse Winding Test (Layer Short, IWT) |
External option |
Channel Programming |
H/L/X in 8CHs |
Withstanding Voltage Test |
Output Voltage |
AC: 0.05~5KV / DC: 0.05~6KV
|
Load Regulation |
±(1% setting + 0.1%full Scale) |
Voltage Regulation |
2V |
Voltage Accuracy |
1% setting + 0.1% full Scale |
Cutoff Current |
AC: 30mA / DC: 10mA
|
Current Resolution |
AC: 1µA / DC: 0.1µA |
Current Accuracy |
1% of reading + 0.5% of range (1% of reading + 5% of total current) |
Current Frequency |
50Hz / 60Hz |
Test Time |
0.1 ~ 999 sec , continue |
Test / Ramp / Fall / Dwell Time |
0.3 ~ 999 sec., continue / 0.1 ~ 999 sec., off / 0.1 ~ 999 sec., off / 0.1 ~ 999 sec., off |
Wave Form |
Sine wave |
Insulation Resistance Test |
Output Voltage |
DC : 0.05 ~ 5kV |
Voltage Resolution |
2V |
Voltage Accuracy |
1% of setting + 0.1% of full range |
IR Range |
1MΩ ~ 50GΩ |
Resistance Resolution |
0.1MΩ |
Resistance Accuracy |
≧1000V |
0.1MΩ ~ 1GΩ : ± (3% of reading + 0.1% of full range) 1GΩ ~ 10GΩ : ± (7% of reading + 2% of full range) 10GΩ ~ 50GΩ : ± (10% of reading + 1% of full range) |
500V~1000V |
0.1MΩ ~ 1GΩ : ± (3% of reading + 0.1% of full range) 1GΩ ~ 10GΩ : ± (7% of reading + 2% of full range) 10GΩ ~ 50GΩ : ± (10% of reading + 1% of full range) |
≦500V |
0.1MΩ ~ 1GΩ : ± 3% of reading + (0.2*500/Vs)% of full scale |
DC Resistance Measurement |
Test Signal |
<DC 10V. < DC 100mA |
Measurement mode |
2 terminals (2W) / 4 terminals(4W) measurement selectable |
Measurement Accuracy (2W/ 4W) |
1Ω (4W only) |
±(0.5% of reading + 0.5% of range) |
10kΩ |
± (2% of reading + 0.5% of range) / ± (0.5% of reading + 0.05% of range) |
100kΩ |
1kΩ |
10kΩ |
100kΩ |
Flashover Detection |
Setting Mode |
Programmable setting |
Detection Current |
AC : 1mA ~ 15mA, DC : 1mA ~ 10mA |
Secure Protection Function |
Fast Output Cut-off |
0.4ms after NG happen |
Ground Fault Interrupt |
0.5mA ±0.25mA AC, ON/OFF |
Panel Operation Lock |
Present password |
Interlock |
YES |
GO/NG Judgment Window |
Indication, Alarm |
GO : Short sound, Green LED; NG : Long sound, Red LED |
Data Hold |
Least tests data memories |
Memory Storage |
50 instrument setups with up to 20 test steps |
Interface |
RS232*1 (Standard), RS232*1 or GPIB & Handler & Temperature interface (Optional). |
General |
Operation Environment |
Temperature : 0°C~45°C, Humidity : 15% to 95% R.H@≦40°C |
Power Consumption |
500VA |
Power Requirements |
90~132Vac or 198~264Vac, 47~66Hz |
Weight |
Approx. 20KG |
Certifications |
|
|
Lưu ý: Nếu một thiết bị nào đó không được liệt kê ở đây, điều đó không có nghĩa rằng chúng tôi không hỗ trợ được bạn về thiết bị đó.
Hãy liên hệ với chúng tôi để biết danh sách đầy đủ về thiết bị mà chúng tôi có thể hỗ trợ và cung cấp.
INO: Bán, Báo giá, tư vấn mua sắm và cung cấp, tư vấn sản phẩm thay thế; tương đương, hướng dẫn sử dụng, giá…VNĐ, …USD info@ino.com.vn | 086 9421303 | 02873000184 | QuadTech Inc-Wound Component EST Scanner (Model:Hybrid 2000).