Vui lòng liên hệ với chúng tôi– chúng tôi sẽ liên hệ lại với khách hàng bằng điện thoại hoặc email.
Lưu ý: Tham khảo ý kiến của nhân viên INO sẽ giúp bạn tiết kiệm được thời gian và chi phí khi cần mua sắm. Với sự tư vấn của chúng tôi, bạn sẽ không gặp khó khăn khi tìm hiểu về đặc tính của sản phẩm cần mua.
Ultrasonic Fatigue Testing System
The USF-2000 implements materials fatigue testing at 20 kHz. Consequently, it is possible to evaluate 109 to 1010 order fatigue strength, where it had been difficult to obtain data before. Also, data for 107 cycles can be obtained in about ten minutes.
[Features]
- Accelerated evaluation of the fatigue life of metals and materials is possible at 20 kHz cycle speed. Even 1010 cycle tests are completed in a mere six days.
- Significant stress can be produced by testing under resonance conditions, and furthermore is economical.
- Constant stress amplitude control is possible via the controller (provided) with the PC.
- Optimal for long-life evaluation of materials and high-speed vibration replication experiments.
- The test system is extremely compact, allowing it to be installed anywhere.
- The Windows software provided can be used for easy testing from design of resonance specimens through to actual testing.
[System Configuration]
Part No. | Name |
346-72207
Without air compressor |
USF-2000 Ultrasonic Fatigue Testing System Tester main unit (including installation table) Ultrasonic resonance system Controller (PC) Ultrasonic test control measurement software Cooling system (air compressor*, air drier, air pipes) |
346-72209 | Displacement measuring instrument (eddy current type displacement sensor, resolution 0.5 µm) |
346-52897-01 | Displacement sensor calibrator (High-Performance Micrometer CDE-25 C1) |
– | Displacement recorder, High-speed data logger (sampling speed 1 MHz min.), Measurement laptop computer |
[Specifications]
Test frequency | 20 kHz ±500 Hz (recommended test range 20 kHz ±30 Hz) * The test frequency is determined by the resonance frequency of the specimen. |
Horn end face amplitude | ±10 to ±50 µm |
Test stress range | Stress when the specimen undergoes ±10 to ±50 µm displacement *The stress value depends on the specimen shape and physical properties. |
Power requirements | 200 V 3-phase: 2 kVA (air compressor) 200 V 1-phase: 3.5 kVA (ultrasonic fatigue testing system) 100 V 1-phase: 1 kVA (including PC, displacement recorder, air drier) |
[Testable Materials/Materials That Cannot Be Tested]
Testable materials | Materials that can resonate at 20 kHz, and that generate little heat when resonating (such as high-strength steel, duralumin, titanium alloys, aluminum) |
Materials that cannot be tested | • Materials that do not resonate at 20 kHz • Materials for which specimen installation is difficult • Materials that generate significant heat due to hysteresis energy when resonating at 20 kHz (such as plastics, ceramics) |
Lưu ý: Nếu một thiết bị nào đó không được liệt kê ở đây, điều đó không có nghĩa rằng chúng tôi không hỗ trợ được bạn về thiết bị đó. Hãy liên hệ với chúng tôi để biết danh sách đầy đủ về thiết bị mà chúng tôi có thể hỗ trợ và cung cấp.
INO: Bán, Báo giá, tư vấn mua sắm và cung cấp, tư vấn sản phẩm thay thế; tương đương, hướng dẫn sử dụng, giá…VNĐ, …USD info@ino.com.vn | 086 9421303 | 02873000184 | Shimadzu-Ultrasonic Fatigue Testing System (Model:USF-2000).
Comments are closed.